Application of balloon-type endotracheal intubation fixator in patients with oral endotracheal intubation |
HE Liang WU Jie-hua WANG Meng-meng CAO Xue-qin ZHANG Xiao-xue |
HE Liang WU Jie-hua WANG Meng-meng CAO Xue-qin ZHANG Xiao-xue |
|
|
Abstract Objective To investigate the application effect of balloon-type endotracheal intubation fixator in patients with oral endotracheal intubation.Methods A total of 81 patients with oral endotracheal intubation admitted to the Second Affiliated Hospital of Nanchang University from February 2017 to May 2019 were selected as the research objects.According to the random number table method,they were divided into the control group(40 cases)and the observation group(41 cases).The control group used traditional tape to fix the catheter,and the observation group used balloon-type endotracheal intubation fixator to fix the catheter.The incidence of catheter displacement,intubation fixation time,replacement time of fixation and tape,visual analogue scale(VAS)score,and total incidence of skin complications were compared between the two groups.Results The incidence of catheter displacement in the observation group was 36.6%,which was lower than 62.5% in the control group,the difference was statistically significant(P<0.05).The intubation fixation time and replacement time of fixation and tape in the observation group were shorter than those in the control group,and the differences were statistically significant(P<0.05).The VAS score of the observation group was(4.1±0.9)points,which was lower than the(6.3±1.2)points of the control group,and the difference was statistically significant(P<0.05).The total incidence of skin complications in the observation group was 7.3%,which was lower than 27.5% in the control group,and the difference was statistically significant(P<0.05).Conclusion The application effect of balloon-type endotracheal intubation fixator in patients with oral endotracheal intubation is significant and can be popularized in clinic.
|
|
|
|
|
[8] |
王红,景新华,柏慧华,等.口含嘴式牙垫在经口气管插管患者中的应用[J].中华现代护理杂志,2017,23(14):1916-1918.
|
[9] |
刘佩,吴荷玉,廖丽华.改良经口气管插管固定方法在头面部手术中的应用[J].护理学杂志,2018,33(22):44-45.
|
[10] |
陈映红,梁春媛,张梅芬,等.改良式3M弹力胶带固定法在重症监护病房经口气管插管患者中的应用效果[J].中国当代医药,2020,27(16):240-243.
|
[11] |
唐艾琳,宋敏,杨田.巧用3M弹性柔棉宽胶带制作导管固定器[J].当代护士(上旬刊),2018,25(11):191.
|
[12] |
孙凤凤,翟怀香,唐冰.改良胶布加强固定法在ICU患者行双腔颈内中心静脉导管固定中的应用研究[J].中国实用护理杂志,2017,33(33):2620-2623.
|
[13] |
潘世琴.用改良式经口气管插管固定法为ICU患者固定气管插管的临床效果[J].当代医药论丛,2016,14(1):3-6.
|
[1] |
朱良梅,谢文林,金巧萍.一体式气管插管固定器的设计与应用[J].中华现代护理杂志,2016,22(10):1467.
|
[2] |
刘姗姗,娄小平.重症患者经口气管插管三种固定方法的应用效果比较[J].中国民康医学,2019,31(14):18-20.
|
[3] |
倪胜贤,周淳,张慧琴,等.抽吸式牙刷联合氯己定冲洗在经口气管插管病人口腔护理中的应用[J].护理研究,2019,33(5):846-849.
|
[4] |
白玲,梁文婷,许华.经口气管插管两种固定方法在ICU俯卧位通气患者中的应用价值分析[J].实用临床护理学电子杂志,2018,3(5):130,133.
|
[5] |
刘合顺.喉罩与气管插管对全麻患者血流动力学的影响[J].中国民间疗法,2018,26(14):85-86.
|
[6] |
黎剑云,黎学敏,冼青英,等.吸痰刺激时间对气管插管患者吸痰效果的研究[J].心理医生,2018,24(24):166-167.
|
[7] |
叶小丽.ICU经口气管插管患者非计划性拔管的原因分析及护理效果分析[J].医药前沿,2016,6(8):242-243.
|
[14] |
方秀云,陈腾辉.两种经口气管插管固定方法效果探讨[J].当代护士(中旬刊),2016,23(10):107-109.
|
[15] |
Tomasello G,Damiani F,Cassata G,et al.Simple and fast orotracheal intubation procedure in rats[J].Acta Biomed,2016,87(1):13-15.
|
|
|
|