Investigation study on the reference range of urinary sediment of the healthy population in the local area |
SUN Ying ZHAO Gui-ping SHI Mei-lin PEI Qian-yun |
Clinical Laboratory,Gaochun People′s Hospital of Nanjing City,Nanjing211300,China |
|
|
Abstract Objective To study and determine the reference range of urinary sediment of healthy adults in the region usingⅠQ200 urine sediment analyzer.Methods The normal urine samples from 2225 healthy subjects in our hospital from September 2013 to January 2015 were selected as the research objects,the HT-2000 urine dry chemistry analyzer and urine microscopic examination of two kinds of centrifugal method were used,white blood cells,red blood cells,the transparent tube type was detected byⅠQ200 urine sediment analyzer.The difference between 3 age groups at 21-40,41-60,more than 60 years old and the differences between different genders were compared.Percentile method was used to determine the scope of the three indicators of the reference.Results There was no significant difference in the count of red blood cells,white blood cells,the transparent tube type of different age groups of the male and female(P>0.05).There was significant difference in the count of red blood cells,white blood cells among different age groups of the male and female(P<0.05).There was no significant difference in the count of the transparent tube type among different age groups of the male and female(P>0.05).ⅠQ200 urine sediment analyzer in the 95%confidence interval of the reference range was that the count of red blood cells and white blood cells in male was 0-11/μl,0-12/μl respectively; the count of red blood cells and white blood cells in female was 0-12/μl,0-19/μl respectively.The count of the transparent tube type in the male and female was 0-1/μl.Conclusion The reference range of urinary sediment established in this study is authentic,which can be used in the practical work,and can provide a reliable basis for the diagnosis and curative effect observation of clinical diseases.
|
|
|
|
|
[1] |
顾可梁.尿沉渣管型检查[J].临床检验杂志,2001,19(1):58.
|
[2] |
顾可梁.尿沉渣检查的新进展[J].临床检验杂志,2004,22(5):395.
|
[3] |
谭欣.ⅠQ200与显微镜测尿中小圆上皮细胞对比分析及参考区间的建立[J].临床输血与检验,2015,17(3):252-254.
|
[4] |
CLSⅠC28-A3.Defining,establishing,and verifying reference intervals in the clinical laboratory[Z].3rdedition.2010.
|
[5] |
Sasse EA.How to define and determine reference intervals in the clinical laboratory[J].NCCLS Document,1995,15(4):C28.
|
[6] |
黄平,周云丽.尿有形成分检验的现状和发展趋势[J].国际检验医学杂志,2009,30(11):1095-1096.
|
[7] |
刘欣,赵曙光,白雪,等.尿沉渣定量分析仪与尿干化学分析仪在尿液细胞检测中的联合应用[J].国际检验医学杂志,2015,36(18):2679-2680.
|
[8] |
杜娟,徐佳,郭野,等.全自动尿沉渣分析仪的临床应用评价[J].协和医学杂志,2014,5(3):290-296.
|
[9] |
刘志超.尿沉渣自动化分析研究进展[J].医学检验与临床,2011,22(4):73-74.
|
[10] |
Ben-Ezra J,Bork L,McPherson RA.Evaluation of the Sysmex UF-100 automated urinalysis analyzer[J].Clin Chem,1998,44(1):92-95.
|
[11] |
陶虹,陈云峰,葛亮.ⅠQ-200全自动尿沉渣分析仪有形成分分析的评价[J].国际检验医学杂志,2009,30(1):74-75.
|
[12] |
师伟,张国富.ⅠQ200全自动尿沉渣分析仪检测尿红细胞主要性能评价的探讨[J].中国药物与临床,2014,14(1):127-128.
|
[13] |
裴忠亚,孙涛.ⅠQ200尿沉渣分辨正确率调查[J].医学检验与临床,2010,21(3):128.
|
[14] |
郝秀珍.全自动尿沉渣分析仪ⅠQ200的使用与显微镜法的比较[J].中国卫生产业,2014,11(36):23-24.
|
[15] |
许丽萍,毛菊珍,张莹,等.600名健康成人ⅠQ200全自动尿沉渣分析仪的参考值测定[J].苏州大大学学报(医学版),2008,28(1):121-123.
|
|
|
|