Application of insulation tester in risk management of electrosurgical instruments |
LIU Wei |
Disinfection Supply Center, Jiangxi Provincial People′s Hospital, Jiangxi Province, Nanchang 330006, China |
|
|
Abstract Objective To explore the application effect of insulation tester in risk management of electrosurgical instruments.Methods A retrospective analysis was conducted to select 2458 pieces of electrosurgical instruments recovered from the Disinfection Supply Center of Jiangxi Provincial People′s Hospital from April to June 2020 as the control group, including 1931 pieces of operating pliers, 311 pieces of electric coagulation hooks, 135 pieces of power cables,and 81 pieces of other instruments, a total of 60 staff of clinical departments were randomly selected for investigation.In addition, 2510 pieces of electrosurgical instruments recovered from the Disinfection Supply Center of Jiangxi Provincial People′s Hospital from July to September 2020 after the improvement of insulation detection methods were selected as the observation group, including 1952 pieces of operating pliers, 324 pieces of electric coagulation hooks, 147 pieces of power cables and 87 pieces of others, a total of 60 staff of clinical departments were randomly selected for investigation.The total detection rate of insulation damage of electrosurgical instruments was compared between the two groups,and the total satisfaction of the clinical staff was compared between the two groups.Results The total detection rate of electrical surgical instrument insulation damage in the observation group (16.53%) was higher than that in the control group (5.78%), the difference was statistically significant (P<0.05).The total satisfaction of clinical departments in the observation group (100.00%) was higher than that in the control group (91.67%), and the difference was statistically significant (P<0.05).Conclusion In the risk management of electrosurgical instruments, the insulation tester can effectively improve the detection rate of insulation damage, reduce the insulation safety accidents related to electrosurgical instruments, reduce the potential safety hazards, and improve the satisfaction of departments.
|
|
|
|
|
[1] |
赵海璇,王洁贞,安福丽,等.风险管理在手术室电外科中的安全应用[J].护理实践与研究,2012,9(13):7-8.
|
[2] |
张丽,哈丽旦·阿布力孜,邵丽.失效模式与效应分析在降低手术室电外科设备使用风险发生率中的应用[J].新疆医科大学学报,2016,39(12):1606-1609.
|
[3] |
中华人民共和国国家卫生和计划生育委员会.WS310-1医院消毒供应中心第1 部分:管理规范[S].国家卫生和计划生育委员会,2016.
|
[4] |
耿小平,郭莉.电外科技术的发展与应用[M].北京:人民军医出版社,2015:15-37.
|
[5] |
李毅.术中高频电刀的安全使用[J].饮食保健,2020,7(17):277.
|
[6] |
Tixier F,Garc,on M,Rochefort F,et al.Insulation failure in electrosurgery instrumentation:a prospective evaluation[J].Surg Endosc,2016,30(11):4995-5001.
|
[7] |
Espada M,Munoz R,Nobile BN,et al.Insulation Failures in Robotic and Laparoscopic Instrumentation:a prospective evaluation[J].Am J Obstet Gynecol,2008,20.
|
[8] |
戴世斌,崔蕾,孙忠根.风险管理在医院医疗器械管理中的应用效果及对管理质量的影响研究[J].中国医学装备,2019,16(1):125-128.
|
[9] |
丁晓青,许聃聃.消毒供应中心风险管理实践[J].中国社区医师,2012,14(32):351.
|
[10] |
王婷,赵玲玲,何丹.风险评估管理在医疗器械消毒供应中心的应用[J].中国临床护理,2019,11(3):265-267.
|
[11] |
刘东岩,郭大为,王爽,等.有源腔镜器械的绝缘性能检测及破损原因分析[J].中国医学装备,2018,15(11):54-57.
|
[12] |
中华人民共和国国家卫生和计划生育委员会.WS310-1医院消毒供应中心第2 部分:清洗消毒及灭菌技术操作规范[S].中华人民共和国国家卫生和计划生育委员会,2016.
|
[13] |
刘书英.电外科手术器械绝缘层破损的风险和应对[J].护理研究,2019,33(6):1083-1085.
|
[14] |
袁园,刘作辉.电外科手术器械绝缘性能检测[J].中国医学装备,2018,15(5):123-127.
|
[15] |
王少霞,杨丽珍,吴锦凤.电外科手术器械绝缘检测仪在消毒供应中的应用[J].中国医疗器械信息,2020,26(17):177-179.
|
|
|
|